ELPAC MVA100 SERIES 100 Watt Medical Open Frame Power Supply * * * * * * * * Model Number Medical Approval - EN60601-1 Class I 3rd Edition High Efficiency Hold-up Time >24ms at full load Up to 120W with Forced Air Floating Output EISA, CEC Compliant 5-Year Limited Warranty Class II Available, consult factory Output Voltage Output Current Peak Current 1 Total Regulation 2 Typical Efficiency 3 MVA100012A 12.0V 8.3A 10.A 5% 90% MVA100015A 15.0V 6.7A 8.0A 5% 90% MVA100018A 18.0V 5.6A 6.7A 5% 91% MVA100024A 24.0V 4.2A 5.0A 5% 92% Notes 1) Maximum peak load (120W) lasting 500ms with a maximum 10% duty cycle. 2) Includes initial setting, line regulation, load regulation, and thermal drift. 3) Typical at 115VAC INPUT OUTPUT Input Voltage 85 - 264VAC 100 - 240VAC Nominal Input Frequency 47 - 63Hz Input Current <2.0A rms Inrush Current <37A at 230VAC cold start Zero Load Power Consumption <0.5W Earth Leakage Current (Typical) Touch Leakage Current Output Voltage See Table Total Regulation +/-5% Minimum Load No minimum load required Start-Up Delay <750ms Hold-Up Time >24ms at any input voltage Ripple & Noise <1% pk-pk ** Over Voltage Protection 110 - 135% <125A @ 132VAC @ 60Hz Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable <175A @ 264VAC @ 60Hz Over Current Protection 120 - 180% Short Circuit Protection Shutdown, auto-restart (hiccup mode) <50A @ 132VAC @ 60Hz <80A @ 264VAC @ 60Hz Notes **Ripple and noise measured with 20MHz bandwidth; 10F tantalum capacitor in parallel with a 0.1F ceramic capacitor. Inventus Power 1200 Internationale Parkway Woodridge, IL 60517, USA Inventus Power Technical Support: powersupplies@inventuspower.com ELPAC MVA100 SERIES 100 Watt Medical Open Frame Power Supply General Efficiency Environmental Avg. Efficiency 88.6% @ 115VAC; 89.5% @ 230VAC Operating Temperature 0 - 70C (Full load to 50C, derate linearly to 50% load at 70C) MTBF min. 200,000 hours demonstrated Storage Temperature -40C to +85C Size 4.50" (114mm) x 2.5" (63.5mm) x 1.22" (30.9mm) Relative Humidity 5-95%, non-condensing Cooling Natural Convection Vibration All units production tested to 19.6m/s2 Weight 0.45 lbs (0.20 Kg) Power Density 8.7W/in3 (at 120W) EMC & Safety Emissions FCC class B, CISPR11 class B EN61000-3-2, -3 Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11 cTUVus UL 60601-1 Certified by TUV to the following: CAN/CSA-22.2 No.601.1-M90 CB per IEC60601-1 2nd & 3rd Edition CE marked to LVD Input Configuration Output Configuration Input Pin Assignments Mating Connector AMP p/n 640250-3 (or equivalent) Connector (PSU side) AMP p/n 640445-8 (or equivalent) Pin 1 AC Line Connection on AMP p/n 640445-3 Power Supply (or equivalent) Body Mating Connector AMP p/n 640250-8 (or equivalent) Pin 2 <not assembled> Pin 3 AC neutral Mechanical Drawing Inventus Power 1200 Internationale Parkway Woodridge, IL 60517, USA Output Pin Assignments Pin 1 +V1 Pin 2 +V1 Pin 3 +V1 Pin 4 +V1 Pin 5 Return Pin 6 Return Pin 7 Return Pin 8 Return IP_DS_MVA100_7-29-16_01 inventuspower.com (c)2016 Inventus Power, All rights reserved.