MVA100 -- ICCNexergy Page 1 of 3 MVA100 Series 100 W Medical Open Frame Power Supply * High Efficiency: Level V * Up to 120W with Forced Air * High Power Density 8.7W/in3 * Hold-up Time >24ms at full load * EISA & CEC Compliant * Output Floating * Medical Approval - EN60601-1 Class I Elpac Part Output Output Peak Total Typical Number Voltage Current Current Regulation Efficiency MVA100012A 12.0V 8.3A 10.0A 5% 90% MVA100015A 15.0V 6.7A 8.0A 5% 90% MVA100018A 18.0V 5.6A 6.7A 5% 91% MVA100024A 24.0V 4.2A 5.0A 5% 92% Input Input Voltage 85 - 264VAC 100 - 240VAC Nominal Input Frequency 47 - 63Hz Input Current <2A rms Inrush Current <37A at 230VAC cold start Zero Load Power Consumption <0.5W Patient Leakage Current <50A @ 132VAC @ 60Hz <80A @ 264VAC @ 60Hz Touch Leakage Current <125A @ 132VAC @ 60Hz <175A @ 264VAC @ 60Hz Output http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mva100/ 5/27/2011 MVA100 -- ICCNexergy Page 2 of 3 Output Voltage See Table Total Regulation +/-5% Minimum Load No minimum load required Start-Up Delay <750ms Hold-Up Time >24ms at any input voltage Ripple & Noise <1% pk-pk ** Over Voltage Protection 110-135% Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable Over Current Protetion 120 - 180% Short Circuit Protection shutdown, auto-restart (hiccup mode) General Efficiency Avg Efficiency 88.6% @ 115VAC; 89.5% @ 230VAC MTBF min. 200,000 hours demonstrated Size 4.50 Weight 0.45 lbs (.20 kg) Power Density 8.7W/in3 (at 120W) Environmental Operating Temperature 0 - 70C (Full load to 50C, derate linearly to 50% load at 70C) Storage Temperature -40C to +85C Relative Humidity 5-95%, non-condensing Cooling Natural Convection (100W) or Forced Air (120W) Vibration All units production tested to 19.6m/s2 EMC & Safety Emissions FCC class B, CISPR11 class B EN61000-3-2, -3 Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11 Certified by TUV to the following: cTUVus http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mva100/ 5/27/2011 MVA100 -- ICCNexergy Page 3 of 3 UL 60601-1 CAN/CSA-22.2 No.601.1-M90 CB per IEC60601-1 CE marked to LVD MVA100 | ICCNEXERGY | (c) 2011 | http://iccnexergy.com/elpac/medical-power-supplies/open-frame/mva100/ 5/27/2011